Introduction
We
have developed a system that employs the principles of the
oscillometric method (a highly accepted method for determining
indirect blood measurement) which, in addition to measuring pulse
pressure value and heart rate, also provides an index (ASI) for the
degree of arterial stiffness. ASI possesses enormous significance
for determining the efficacy of various diagnoses and treatments for
hypertension. Since ASI is a new index, its measurement principles
and clinical meaning are not yet fully understood.
In order to clearly ascertain the meaning of ASI, we explain the
fundamental principles of blood pressure measurement in this paper.
Next, we indicate the method and way in which ASI is resolved, and
explain ASIfs meaning. We also introduce fundamental ASI-related
data.
1.
Principles
and Characteristics of Indirect Blood Pressure Measurement Based on
the Oscillometric Method.
When
employing the oscillometric method to measure blood pressure,
pressure is applied to a cuff wound around the arm of the patient.
Blood pressure is then measured from very slight oscillations
generated within the cuff when pressure is applied. As with
auscultation, the arm is wound around the arm and pressure is raised
above systolic pressure. Pressure is then reduced. As pressure is
reduced, the pulsation of the artery varies, and very slight
pressure changes occur in accordance with this arterial
volume-pressure variation. The oscillometric method determines blood
pressure based on these slight pressure variations.
When
cuff pressure is raised above systolic pressure and then reduced,
the pulse wave amplitude suddenly becomes wide near systolic
pressure, and then gradually increases until it reaches its greatest
width near mean pressure. The oscillometric method may be explained
from studying the causes of this pulse wave amplitude variation.
In
order to understand the oscillometric method, it is essential first
to comprehend the relationship between the composition of the
arterial wall and its dynamic characteristics.
1-1
Arterial
Vascular Pressure-Volume Characteristics
The
elastic fibers and collagen fibers of the arterial wall provide the
arterial wall with its extensibility characteristics. Roach and
Burton applied formic acid to the artery to selectively remove its
collagen fibers and then investigated the extensibility
characteristic of the inner membrane and tunica media, which are
mainly composed of elastic fiber and smooth muscle.
The result was that the layers of inner
membrane and tunica media proved to be rich in extensible protein
character fibers, and that since these were bundled together in a
disorderly direction, it was clear that the elasticity coefficient
was low and extensibility was high. Roach and Burton also applied
trypsin to the arteries and blood vessels, removing the inner
membrane and tunica media, and investigated the extensibility of the
outer membranefs collagen fibers. The result was that the collagen
fibers proved high in dynamic strength compared to the elastic
fibers, but their extensibility was strikingly low.
Using these results, we will consider the movement
power (inner pressure) in the arteries and the extensibility of each
of the arterial membranes.
Figure
1 displays the arterial structure and extensibility. As indicated in
the figure, since the outer membrane is not stretched or extended in
regions of relatively low pressure in an artery, arterial wall
extensibility depends primarily on the properties of the inner
membrane and tunica media. The layers of these membranes possess
high extensibility, and the arterial wall indicates large elongation
depending on variation of inner pressure. As a result, in
high-pressure regions, the inner membrane and tunica media extend
adequately and the outer membrane that covers the arterial exterior
also extends. At this time, the extensibility of the whole vascular
segment is decided based upon the properties of the outer membrane,
which has the lowest extensibility. Thus, the arterial
extensibility, corresponding to pressure variation, becomes
strikingly low. Arterial inner pressure increases and volume
variation, in particular differences between the tunica media and
outer membrane's elastic modulus and manners of working, may be
explained in terms of springs. We will describe the tunica media
characteristic as a soft or limp spring and the outer membrane
characteristic as a hard spring. Since force moves in the soft
spring (tunica media) when the inner pressure is in a low state, elongation or stretching
becomes larger with variations in inner pressure. When the inner
pressure increases even further, arterial volume increases, and
after elongation or stretching proceeds beyond a certain point, the
hard spring (outer membrane) will respond and the arterial
extensibility will decrease.
As displayed in
Figure 2, the property of a normal
artery manifests as the result of the synthesis between the
properties of the tunica media and the outer membrane, and variation
in vascular extensibility depends upon the rise and fall of inner
pressure.
In this manner, since the three membranes composing the
arterial wall all possess different extensibility properties, the
arterial pressure volume property depends on the inner pressure (or
on the volume) and is unique. In Figure
3, the relationship between
the arterial pressure-volume property and the volume pulse wave is
modeled. The vertical axis represents pressure, and the horizontal
axis represents arterial vascular volume. As shown in the same
figure, the arterial pressure-volume property indicates a strong
non-linearity, and does not indicate volume variation, which is in
proportion to pressure variation. Thus, even in the case that
arterial pressure is equal, if the inner pressure acting within the
blood vessel differs, a change in the size of the volume variation
corresponding with that pressure will be generated in regards to the
arterial pressure.
1-2
Principles
of Blood Pressure Determination
Figure
4 represents the blood pressure added to the artery during blood
pressure measurement. For measuring blood pressure, a cuff is wound
around the arm and compressed with air. At this time, the artery is
compressed as a result of cuff pressure. The pressure added from
outside the artery by the cuff occurs in addition to the arterial
inner pressure (the pressure produced by pulsation of the heart ,
i.e. the blood pressure inside the arterial wall).
In
this situation, the "pressure" of the pressure-volume property
vertical axis in Figure
5, which displays pulse wave variation
produced during decompression, may be considered as the difference
between the inner and outer arterial pressure (transmural pressure),
in other words, the difference between blood pressure and cuff
pressure. When the cuff pressure displayed in the figure is 0 during
measurement, the only pressure active in the artery is the blood
pressure, and the arterial volume variation produced by this
pressure appears as 7 in the figure. When the cuff pressure is a
larger value than systolic pressure (difference between arterial
inner and outer pressure is negative), the volume variation appears
as 1 in the figure. When gradually reducing cuff pressure from a
pressure above systolic pressure, arterial volume varies with cuff
decompression and the amplitude varies as reflected in 1 to 7. In
other words, as pressure decreases, the artery expands and the
amount of arterial pressure-volume variation corresponding with
pulse pressure gradually becomes larger. When the mean arterial
inner-outer pressure is equal, the volume variation corresponding
with the pressure variation (pulse pressure) reaches its largest
point as indicated in 2 in the figure. As a result, the cuff
pressure where the pulse wave amplitude reaches its highest point is
identical with mean pressure.
When
cuff pressure reaches the mean blood pressure area, the inner and
outer arterial pressure is on average 0. In this area, since the
extensibilities of the inner media and tunica media are at their
greatest point, the arterial pulsation, in other words, the volume
variation, also reaches its greatest point at this time. Since
elastic modulus increases simultaneously with swelling of the artery
when cuff pressure is further reduced, the pulse wave amplitude
decreases. When cuff pressure falls below diastolic pressure, the
artery becomes even stiffer, and the pulse wave amplitude rapidly
becomes small. This is due both to the fact that the outer membrane
characteristic manifests in regards to arterial swelling and that
the elastic modulus of the collagen fibers forming the main
component of the outer membrane causes the artery to become
stiff.
Since
there are limits in the extent to which the cuff can effectively be
compressed during actual measurement, the oscillometric method
cannot correctly measure systolic pressure by depending upon the
existence or non-existence of the pulse wave. There is also no
sufficient explanation of a logical basis for determining diastolic
pressure. In determining systolic and diastolic pressure, an
analysis program has been prepared to improve concordance with blood
pressure values derived from the auscultation method. The result is
that that the blood pressure values of oscillometric and
auscultation match relatively well on a statistical basis. With the
CardioVision, cuff compression and decompression are conducted
automatically, and in this process the pulse wave amplitude
variation is analyzed and blood pressure is determined. Figure 6
displays the relationship between the cuff pressure and arterial
amplitude that appear during the indirect measurement of blood
pressure using the oscillometric method. In the oscillometric
method, systolic pressure is determined as the pressure
corresponding with the increased point of pulse wave amplitude, mean
blood pressure as the largest point of amplitude, and diastolic as a
sudden decrease in amplitude.
Thus,
the computer becomes essential for measurement with the
oscillometric method and for pattern analysis. Since the
oscillometric method includes an automatic analysis program, it has
the following merits.
1.
Subjectivity of the measurer is not a factor in accuracy of
measurement.
2.
Unlike the auscultation method where a microphone is used, the
operation of precise application to the upper arm position is
unnecessary and is also effective against noise.
For
these reasons, the oscillometric method is being increasingly used
for infants and newborns and in households as well. In the clinical
field, it is finding increasing use as an easy-to-use blood pressure
monitor. Since
arterial dynamic properties manifest in measurement data collected
from blood pressure measurement using the oscillometric method in
particular, it is possible to offer data regarding the properties of
the arteries in addition to the blood pressure. In this research,
particular attention was paid to this point in detecting arterial pulse
wave amplitude. We then evaluated the mechanical properties of the
artery derived from the detected pulse wave, and thus were able to
calculate the degree of arterial wall stiffness. If this method is
employed in the clinical field, we expect that it will prove
effective not only in blood pressure measurement but also in
screening for arterial stiffness, a common disease.
1-3 Why Is
Arterial Volume Variation Acquired from Cuff Pressure
Variation?
In blood pressure
measurement using the oscillometric method, the arterial volume
under cuff pressure is measured as slight variations in cuff inner
pressure. In fundamental explanations of blood pressure measurement,
the relationship between blood pressure and arterial volume
variation is employed in the determination of blood pressure. For this
reason, it is essential to precisely confirm to what degree cuff
inner pressure variation corresponds with arterial volume
variation.
We will now
consider measurement of blood pressure on the upper arm. In the area
underneath the cuff, we assume that an arterial volume (load)
variation V is
generated from one cardiac output (stroke volume). Since the tissue
surrounding the artery may be considered to possess incompressible characteristics, the arm volume variation equals the
arterial pressure variation. The cuff covering the circumference of
the arm is inflated with air. Since the edges of the cuff are
restricted with fabric that is not extendable, the arm volume
variation is identical to the volume variation of the air in the
cuff.
Now, when only
V of
the arm volume variation increases, the cuff indicates a
|V volume variation. During blood pressure measurement, V is
the amount of air in the cuff and thus is the average pressure. In a
constant temperature situation
P~Vconstantk
is established from
a gas state equation.
Since
|V volume variation in regards to the cuff is provided by
arterial volume variation generated from cardiac pulsation, volume
variation P is
generated in accordance with this volume variation. At this time,
since the equation
io{oj~iu|uj
is established, if
P~V is
ignored because it is miniscule in comparison with other terms,
the equation becomes
P~V{V~P|P~V
and when
P~Vis
substituted, the
equation becomes
V~P|P~VO
Thus, the volume
variation V is
expressed as
VP~V^P
After the cuff was applied experimentally and the
relationship between the cuff inner pressure and the air volume was
investigated, it proved that when pressure was reduced
from systolic pressure to the diastolic pressure area,
P^V
increased gradually from systolic pressure value (the beginning
value), but variation in this range remained within 20% and even if
one considers this as a constant value, one may conclude that
precision measurement of volume variation has no effect on blood
pressure determination. Moreover, P^V
value variation was under 5% for mean pressure area cuff pressure
variation of approximately 20mmHg. The ASI standard value ranged from
approximately 30 to 80, corresponding with cuff pressures of 3 to 8
mmHg. In this case, error was further reduced. When ASI is computed
with this method, it is safe to assume that no significant problem
arises in computing volume change from slight variations of cuff
pressure.
2.
Classification and Evaluation of Circulation Dynamics Using
CardioVision
In general, since
the pulse wave is determined from the relationship between the pulse
pressure and the arterial pressure-volume characteristic, the pulse
wave amplitude variation changes with the pulse pressure or the
arterial elasticity property variation. The pulse pressure is
determined from one stroke volume and the arterial elastic
modulus.
On the other hand,
since the arterial pressure-volume property is influenced by
circulation dynamics during measurement, the pulse wave varies
specifically according to various diseases. In the oscillometric
method, since blood pressure is determined as the cuff is gradually
decompressed, the pulse wave amplitude pattern manifests from the
arterial volume variation (elastic modulus variation) causing
variation in the difference in arterial inner and outer pressure
(transmural pressure) and the pulse pressure. Since these parameters
vary according to different types of circulatory disease and
circulatory dynamics, the pulse wave pattern also varies. Until now,
pattern variation was thought to be only an error factor (the causes
of which were unknown) in precision blood pressure
determination.
CardioVision analyzes the relationship between the pulse wave
amplitude pattern appearing during blood pressure measurement
employing the oscillometric method and circulatory dynamics, and
then classifies the results into the 5 basic patterns A to E
displayed in Figure
7.
A record obtained
from a normal circulatory system displays as A (see Figure
8). When
arterial stiffness is normal, the pulse wave during cuff
decompression appears at the point at which cuff pressure approaches
systolic pressure. It increases after this point, and after reaching
its greatest amplitude, gradually decreases. The size of pulse wave
amplitude is determined by arterial extensibility and the size of a
single cardiac output. A pulse wave amplitude which extends to its
greatest length after systolic pressure is reached and then
subsequently decreases appears as a mountain shape with a single
peak. This is classified as Pattern A.
Pattern B appears
in case of low blood pressure, anemia, and shock. Arterial
extensibility is normal, but since cardiac output is reduced, pulse
pressure and pulse wave volume variation become small in accordance
with falling blood pressure. This case is indicated with a
mountain-shaped pattern of low amplitude.
Pattern C appears
in the case of arterial stiffness, diabetes, obesity, old age, and
high stress. These diseases are said in general to lower arterial
elasticity. Pattern C appears because the mechanical properties of
arteries in these cases differ from normal arteries. As indicated in
Figure 9, when the arterial has undergone stiffening, this
stiffening may be considered to have begun originally when damage
occurs in part of the arterial inner membrane and tunica media
containing smooth muscle and collagen fibers rich in extensibility.
The impaired areas undergo thickening of the arterial wall and
collagen formation (adhesion) and then subsequently become stiff. In
comparison with normal or healthy arteries, the pressure-volume
variation displays in particular a sharp linear line from the point
at which the arterial inner and outer pressure difference is 0. When
this is measured with CardioVision, the resulting arterial volume
variation displays a wide flat pattern where the cuff pressure and
mean pressure are either equal or near equal (see Figure
10). The
extent of arterial volume variation becomes smaller in regards to
constant pulse pressure due to lowered extensibility, but when the
artery is stiff, or when strong stress is added to the circulatory
system, since blood pressure rises and pulse pressure becomes large
due to increasing vascular resistance, lowering of arterial
extensibility does not necessarily cause the size of the pulse wave
amplitude to fall. In CardioVision, with the goal of evaluating
Pattern C, a numerical value is calculated to correspond with the
width of the flat part of the pulse wave, which is the pattern's
characteristic. Since this value increases with the stiffness of the
artery, it is possible from this value to present the Arterial
Stiffness Index (ASI) as numerical number information.
Pattern D is
revealed in the case of arrhythmia. With arrhythmia, since the
heart's diastole is not constant, one-time cardiac output becomes
irregular. As a result, since the blood pressure value and pulse
pressure from one stroke volume changes to a large degree,
measurement of blood pressure becomes difficult. Moreover, since the
arterial volume variation is not constant, both the pulse wave
amplitude and the pattern displayed from the pulse wave width become
irregular (see Figure
11).
Pattern E appears
in cases of cardiac disease. In many types of cardiac disease, low
cardiac output is maintained, and if this situation is prolonged,
circulatory reflexes occur to maintain steady blood pressure. Among
these, smooth muscle stress occurs in many arteries, and the
conditions become different from those for a normal arterial
pressure-volume characteristic (see Figure
12).
3. Principles of Arterial Stiffness Index (ASI) Measurement
Using CardioVision
3-1 The Meaning
of ASI
Figure 13 indicates
a comparison of a normal artery pattern and a sclerotic artery
pattern in regards to arterial pressure-volume characteristic. As
the figure indicates, in the area where arterial inner pressure is
low due to a stiffened tunica media, a steeply increasing linear
slope indicates the pressure-volume relationship. As a result, the
bent line for the whole arterial segment pressure-volume
characteristic moves in an upward direction. The figure's vertical
axis may be substituted for blood pressure-cuff pressure in arteries
undergoing pressure added from the cuff. Thus, with the pulse
pressure under a constant state, the pulse wave amplitude during
cuff decompression manifests as this pattern. In the same figure, in
the case of a tunica media with normal elasticity, the pulse wave
pattern indicates the previous peaked mountain shape (a in Figure
13), but in a sclerotic artery, the pulse wave amplitude pattern
shape becomes trapezoidal (b in same figure). The upper part of the trapezoid width is
equivalent to the range or width when in the process of cuff
decompression the difference between blood pressure and cuff
pressure (transmural pressure) passes the pressure part
corresponding with the linear part of the tunica media.
Since this range or
width indicates time in pulse wave sequences obtained in actual
measurement, the range seen in measured patterns varies with the
speed of cuff decompression. If one, rather than considering the
upper part of the trapezoid as time, considers the flat area as cuff
pressure variation corresponding with this range or width, the flat
region may be expressed as information truly independent of
decompression velocity. This range or width is expressed in units of
pressure and corresponds with the arterial pressure-volume
characteristic that depends on the elasticity of the tunica media.
Thus, if the tunica media is soft, the pressure range or width
corresponding with the linear part becomes small, and in the reverse
situation, if the tunica media is stiff, this range becomes wide. If
we view the tunica media pressure-volume characteristic
as a linear (straight line), the pressure range or width is in
direct proportion with the elastic modulus.
When the cuff
pressure variation part corresponding with the trapezoidal upper
part, (which is recognized in pulse wave amplitude example patterns)
is detected, the pressure range or width becomes a value in direct
proportion to the elastic modulus of the arterial tunica media
level. In this research, this pressure value was applied as an index
from which arterial stiffness can be determined by designating the
arterial stiffness index as
ASI=cuff pressure
range (corresponds with pulse wave pattern trapezoidal part)
X10 Here, the reasons for providing a multiple of 10 for cuff pressure
range are: (1) the value obtained is a very small
in terms of pressure (mmHg); (2) one can express this value as an
integer when using this index in a clinical situation; (3) while it
is true that ASI value for gnormalh is two digits (below 80), a 3
digit value is necessary if the index is to express scores of 100 or
greater for arterial effect, and moreover, makes it easier to
determine numerically the meaning of "stiff".
3-2
Fundamentals of
ASI (Arterial Stiffness Index) Calculation
Figure 14 displays
the relationship between ASI and the arterial pressure-volume
characteristic. The upper part of the figure displays the arterial
pressure-volume characteristic while the lower part displays the
relationship between the cuff pressure during actual blood pressure
measurement and the pulse wave pattern. From the figure, the ASI is
understood to be in direct proportion with the tunica media elastic
modulus, which falls within the arterial elastic modulus.
When actual
measurements are conducted based on this method, the pulse wave
sequences do not necessarily form a perfect trapezoidal shape but
frequently exhibit a gentle curved shape. This is because the tunica
media characteristic displayed here does not exhibit a linear line
and also because of the manifestation of combined characteristics
with the outer membrane. For this reason, in calculating the pulse
wave trapezoidal part in this research, the pattern's highest value
was confirmed as 100% and the area from that point to the point
where the pulse wave amplitude falls to 80% as the flat part. Within
this range, a detailed detection of the flat part is conducted. The
areas between pulse waves are filled in with imaginary pulse wave
sequences, and after applying noise correction smoothing using
mobile addition averaging, the flat region is calculated as the area
where the pulse wave height variation is within
5%.
One can also
explain the meaning of ASI using a different expression. In Figure
15 the arterial pressure-volume characteristics of a normal artery
and sclerotic artery are indicated on the left-hand side. A bent
line with an incline indicates arterial compliance. Arterial
compliance expresses arterial softness or extensibility. The right
hand graph in the figure indicates the change of compliance in
arterial inner pressure with regards to the two types of arteries
displayed on the left-hand side of the figure. In this case, the
compliance characteristic of a soft artery becomes suddenly and
characteristically large near an inner pressure of 0. On the other
hand, in the sclerotic artery, the compliance value remains high in
comparison with the normal artery from 0 across a fairly high inner
pressure range. However, the compliance value for the sclerotic
artery does not become all that large nor does the
compliance-pressure value vary that much in a high compliance value
region, rather it becomes a flat-shaped graph. Figure 16 explains
the dynamic meaning of the artery as indicated by ASI, measured
using CardioVision. To account for dispersion due to blood pressure variation
arising from breathing and other physiological factors, the
reliability of ASI as an index is further improved by multiple
measurement and averaging.
4.
Points of Difference
Between ASI and Other Arterial Stiffness Indexes
4-1 Pulse Wave
Velocity
The Pulse Wave
Velocity indicates the velocity of the pulse wave transmitted
through the artery. The arterial pulse wave velocity is expressed in
the following manner.
PWVγidEh^ΟEcj
@@@@@@@darterial
elastic modulus iYoung's Modulus)
@@@@@@@arterial
wall
thickness
@@@@@@@Οblood
density
@@@@@@@carterial
diameter
Here, a large E
corresponds with a sclerotic artery. Οis blood density, but may be considered to be constant
within the circulatory system. Moreover, in considering an analogous
shape for the artery, since h/D becomes constant, differences in
arterial thickness arising from differences in physique and other
factors are unrelated to the PWV.
Thus, in the case
of arterial stiffness, since E and h become simultaneously big, PWV
also increases. The PWV is measured clinically from the distance and
time between a heartbeat and arm and peripheral artery pulse wave
emission (see Figure
17). Moreover, since PWV is expressed as the
square root of variation in several factors, the rate of that
variation becomes small in relation to the original values. The fact
that PWV is related to age also suggests a relationship with
arterial wall stiffness. In addition, it is said that PWV can detect
hypertension and diabetes early stage angiopathy.
However, when this
index is applied clinically, sufficient care is necessary during
blood pressure measurement. As explained previously, the dynamic
characteristic of the artery is dependent on arterial inner
pressure. Since arterial stiffness varies with the degree of
arterial swelling, even if the blood pressure is at a normal level,
the artery may swell to the point at which it causes the outer
membrane to stretch or expand when a load of inner pressure is at
work inside the artery. The artery's Young Modulus E becomes large
in a non-linear fashion with the increase in arterial inner
pressure, and thus, PWV varies with pressure. For these reasons, if
the blood pressure varies when measuring the same person with the
same artery, PWV will vary. In particular, since
medications for reducing arterial stiffness and blood pressure are
used in combination when treating arterial stiffness, it becomes
essential to determine which of these medications causes a reduction
in pulse wave velocity. Thus, PWV cannot be employed as an index to
directly compare measurement values.
4-2
Compliance
Compliance (C) is
the proportion (relationship) of arterial volume variation
(’V) and pressure variation (’P)
and thus is defined as
b’u^’o
As explained
previously, when the compliance value is large, the artery is soft.
As is clear from the arterial pressure-volume characteristic,
arterial compliance is large when blood pressure is low and low when
blood pressure is high. The compliance value varies according to the
extent to which the arterial wall is being stretched. In other
words, compliance varies with arterial inner pressure (blood
pressure). For this reason, when expressing arterial stiffness with
compliance, it is impossible to determine which artery is stiffening
by comparing compliance alone.
Moreover, since the
size of the artery undergoing measurement influences compliance, an
even greater arterial pressure-volume variation is indicated in a
large artery of a large subject in response to the same pressure
variation; consequently, the compliance measurement value increases
with the thickness of the artery. Thus, comparing compliance in arms
of different size in thickness becomes a problem.
It is not easy to
directly measure compliance in actual measurement of living
organisms. Instead, the pulse pressure may be employed as an easily
measured index. The pulse pressure is an index related to
circulatory system compliance. For instance, if one assumes that the
blood volume ’u transmitted through the circulatory system from one
stroke volume is constant, the arterial inner pressure variation
’o that appears as pulse pressure may be expressed as
’o’u^b
From this equation,
it may be understood that ’o is
in indirect proportion to compliance, and in the same way that
compliance is dependent on blood pressure at the time of blood
pressure (mean blood pressure) measurement, the pulse pressure is
also an index related to mean pressure during
measurement.
@
|